| Model Number |
201-02-35 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pending device return and investigation.
|
| |
|
Event or Problem Description
|
|
As reported, during trialing with the 35mm patella trial, while removing the patella trial one of the patella pegs broke off inside the patient.The surgeon was able to remove the nub using a drill.This led to a surgical delay of about 5 minutes.The patient was last known to be in stable condition following the event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the broken instrument reported was likely the result of numerous surgical uses and subsequent sterilization cycles over a 17-year timespan, which resulted in a fractured peg during removal.However, this cannot be confirmed as the device was not returned for evaluation.
|
| |
|