• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U357564
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a video was provided for review.The investigation of the reported event is currently underway.Expiry date: 02/2025.
 
Event or Problem Description
It was reported that during an angioplasty procedure, leak was allegedly felt while inflating the pta balloon.It was further reported that the pta balloon allegedly had a hole at the end when tested outside the patient.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one ultraverse 035 pta dilatation catheter was received for evaluation.During visual evaluation, no kinks were noted to the catheter shaft and no anomalies to the luers/y-body.During functional evaluation, an in-house presto inflation device was used to inflate the balloon.During the inflation, and under microscopic evaluation, a pinhole rupture was noted to the proximal end of the balloon.No other functional testing was performed.One video was provided and reviewed.The video showed the ultraverse 035 balloon being inflated.Liquid was seen streaming from the proximal end of the balloon, but the exact source of the leak could not be determined.No other specific anomalies were noted.Therefore, the investigation is confirmed for the reported pinhole balloon rupture as a pinhole rupture was noted to the balloon under microscopic evaluation of the returned sample, and thus the source of the liquid seen leaking from the balloon from the provided video is determined to be from the pinhole rupture.A definitive root cause for the alleged pinhole balloon rupture could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 02/2025).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly leaked upon inflation.It was further reported that the balloon allegedly had a hole at the end.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15884447
Report Number2020394-2022-00946
Device Sequence Number5284107
Product Code LIT
UDI-Device Identifier00801741092954
UDI-Public(01)00801741092954
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K142261
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberU357564
Device Catalogue NumberU357564
Device Lot NumberCMGQ0603
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/04/2022
Supplement Date Received by Manufacturer01/11/2023
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age75 YR