| Model Number |
A35HPV10040080 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an attempt was made to use fortrex balloon to treat a lesion with central venous stenosis.It was reported that when the balloon was inserted into the patient during the procedure, it was found that the mid-end of the balloon ruptured.The balloon ruptured before reaching the burst pressure.The procedure was completed by replacing the balloon.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis.The device was returned with the balloon in a pre-inflated state with the balloon folds intact a 20ml water filled syringe was used to flush the device and a steady stream of water was observed exiting the tip a 0.035¿ guidewire was loaded through the tip and exited the inflation port of the hub without any difficulty an indeflator with pressure gauge was to inflate the balloon to its nominal pressure of 8atms and to its rated burst pressure (rbp) of 16 atms.There was no issue inflating the balloon and the balloon was able to hold pressure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|