Model Number 0286
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Codes Electric Shock (2554); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that this right ventricular (rv) lead showed a gradual increase in the shock impedance into high, out of range (oor) values.A delivered shock registered a normal shock impedance, but the daily measurements keep showing oor values.A commanded shock was suggested to test ability to charge and deliver without an open circuit fault.The rv lead remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that this right ventricular (rv) lead showed a gradual increase in the shock impedance into high, out of range (oor) values.A delivered shock registered a normal shock impedance, but the daily measurements keep showing oor values.A commanded shock, defibrillation threshold (dft) test was suggested to test ability to charge and deliver without an open circuit fault.The patient underwent the dft procedure.The device converted ventricular fibrillation, despite the high shock impedance.So, no further action was taken.The rv lead remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE G
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15884898
Report Number2124215-2022-49892
Device Sequence Number5071906
Product Code NVY
UDI-Device Identifier00802526431418
UDI-Public00802526431418
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P910073/S077
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/27/2016
Device Model Number0286
Device Catalogue Number0286
Device Lot Number313506
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer11/23/2022
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/27/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age69 YR
Patient SexMale
-
-