| Model Number |
PTH |
| Medical Device Problem Codes |
High Test Results (2457); Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated they received questionable results for four samples collected from the same patient and tested with the elecsys pth immunoassay on a cobas 8000 e 801 module (serial number (b)(4)).The results did not agree with the patient's clinical status.It was asked, but it is not known if any incorrect results were reported outside of the laboratory.The first sample resulted in a pth value of 462 pg/ml when tested on the e 801 analyzer on (b)(6) 2022.The second sample resulted in a pth value of 440 pg/ml when tested on the e 801 analyzer on (b)(6) 2022.The third sample resulted in a pth value of 415 pg/ml when tested on the e 801 analyzer on (b)(6) 2022.This sample was treated with polyethylene glycol (peg) and repeated on the e 801 analyzer on (b)(6) 2022, resulting in a raw value of 17.5 pg/ml, which calculated to a final value of 35 pg/ml (8.4 % recovery) when accounting for the dilution factor.The fourth sample resulted in a pth value of 404 pg/ml when tested on the e 801 analyzer on (b)(6) 2022.This sample was repeated on an architect analyzer, resulting in a pth value of 895.0 pg/ml on (b)(6) 2022.The patient had no clinically confirmed pathology related to the parathyroids.The patient has normal calcium and the parathyroids are normal in imaging studies.It is expected for the patient to have a normal pth value between 15 and 65 pg/ml.
|
| |
|
Additional Manufacturer Narrative
|
|
The sample results were reported outside of the laboratory where they were questioned by a clinician since they did not match the patient's clinical status.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample from the patient was provided for investigation.Initial investigations of the sample determined that it contains heterophilic antibodies which interfere with the assay.The specific type of interference is still trying to be determined.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
Quality controls were acceptable.There was no indication of an instrument related issue.Further investigation of the patient sample determined it contains a heterophilic antibody interferent and an additional unknown interferent.No further investigations could be performed on the sample as the sample material was used up during testing.Additional sample material was provided by the customer for investigations.For investigation of this material, the sample was fractionated by size exclusion chromatography.Reactivity with elecysys pth and pth (1-84) kits is only found in fractions where there is a molecular weight of > 150 kda, which may correspond to igg bound to pth.The results show an autoantibody-based interfering factor in the patient's blood leading to a higher signal and consequently falsely elevated pth.The investigation could not identify a product problem.
|
| |
|