| Model Number |
132-32-51 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Synovitis (2094)
|
| Date of Event |
11/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, approximately 4.5 years post op the initial tha, this 84 y/o female patient was revised due to pain.Surgeon found synovitis and wear on poly.Patient last known to be in stable condition.Explants unavailable.No other information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: pending evaluation.Concomitant medical products: (b)(4), 142-32-07 - cocr fem head 32mm +7 offset 12/14.(b)(4), 186-01-50 - integrip cc, cluster 50mm, g1.(b)(4), 188-00-05 - wedge plasma s/o sz 5.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) based on the available information, the patient involved meets the following risk criteria for early prosthesis wear/osteolysis: significant edge loading of the femoral head on the acetabular liner and combination of largest available femoral head and/or thinnest available acetabular liner was used.The prosthesis wear/osteolysis was able to be confirmed from the provided radiograph/explanted devices.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, h6: clinical codes, component code, type of investigation and investigation conclusion.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|