• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL CMEMS PRINTER SERIAL CABLE; SYSTEM, HEMODYNAMIC, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL CMEMS PRINTER SERIAL CABLE; SYSTEM, HEMODYNAMIC, IMPLANTABLE Back to Search Results
Model Number CM3011
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One i2 cardiomems hospital electronics system was returned for investigation.Inspection of unit revealed exposed wires on the printer serial cable due to physical damage.A replacement serial printer cable was used and the unit was able to conduct successful diagnostic testing.The root cause of the exposed wires was isolated to physical damage due to an undetermined event.
 
Event or Problem Description
This report is to advise of an event observed during analysis confirming exposed wires of the printer serial cable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CMEMS PRINTER SERIAL CABLE
Common Device Name
SYSTEM, HEMODYNAMIC, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL
387 technology circle nw suite 500
atlanta GA 30313
Manufacturer (Section G)
ST. JUDE MEDICAL
387 technology circle nw suite 500
atlanta GA 30313
Manufacturer Contact
bob fryc
one st. jude medical drive
st. paul, MN 55117
6517565400
MDR Report Key15885365
Report Number3004936110-2022-00692
Device Sequence Number12086818
Product Code MOM
UDI-Device Identifier05415067029331
UDI-Public05415067029331
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P100045
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCM3011
Device Catalogue NumberCM3000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2022
Initial Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/05/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-