• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number MI1000 MED-EL CONCERT PIN
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Event or Problem Description
Initial information had been received that further medical intervention is considered.It is unknown if there had been a change in the recipients hearing performance with the device or if an accident or trauma occurred.The recipient has been reimplanted on (b)(6) 2021.
 
Additional Manufacturer Narrative
Despite several requests, no information has been received from the field about the circumstances which led to explantation.However, device investigation revealed damage to the active electrode which is consistent with minute device mobility leading to a device failure over time due to fatigue wire breakages.This is a combined initial and final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key15885521
Report Number9710014-2022-00927
Device Sequence Number20162679
Product Code MCM
UDI-Device Identifier09008737062750
UDI-Public(01)09008737062750
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/14/2018
Device Model NumberMI1000 MED-EL CONCERT PIN
Device Catalogue Number07680
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2022
Initial Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/14/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age8 YR
Patient SexFemale
-
-