Model Number 10143406
Medical Device Problem Code Computer Software Problem (1112)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2022, siemens became aware of a malfunction that occurred while operating the arcadis varis gen 2 system.On november 16, 2022 information was received that the issue occurred during an emergency procedure when the user started fluoro, however, several seconds later the software froze.The user restarted the system, but the issue recurred.The procedure was completed on an alternate unit.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported event.A supplement report will be filed if additional information becomes available.Internal id #: (b)(4).
 
Additional Manufacturer Narrative
H10: siemens completed the investigation of the reported event.The investigation was performed based on expert discussions considering the complaint description, customer service reports, and system history.According to available information, the software was frozen at the start of the system use.As a result, the procedure was continued and finished using an alternative system.During the onsite service intervention, the system started normally, and the described error scenario could not be reproduced.However, the error codes submitted confirmed that there were problems with database access and software.Based these findings, the inside of the pc was cleaned, the internal hardware (e.G., ram, boards) was checked, the bios battery was preventively exchanged, and the software was completely restored.The error mentioned in the complaint has not recurred and has not been reported since the onsite visit.The occurrence rate of the aforementioned error pattern was checked.A possible error accumulation or even a systematic error, which leads to a corrective action of the installed base, could not be determined by the investigation.The incident described was not classified as a reportable adverse event after detailed investigation.Neither serious injury, death, nor an unexpected or prolonged hospitalization of the patient or any other person occurred or could be expected.The corresponding probability according to the risk analysis is in the range of inconceivable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCADIS VARIC GEN2
Common Device Name
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1 or
rittigfeld 1
forchheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1 or rittigfeld 1
rittigfeld 1
forchheim, 91301
GM   91301
Manufacturer Contact
rebecca tudor
40 liberty blvd.
65-1a
malvern, PA 19355
4843234198
MDR Report Key15885624
Report Number3004977335-2022-52158
Device Sequence Number5298453
Product Code OXO
UDI-Device Identifier04056869009018
UDI-Public04056869009018
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K051133
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10143406
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2022
Supplement Date Received by Manufacturer01/12/2023
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-