AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE W/STYLET W/ ENFIT CONNECTOR; DH CPK NG TUBES Back to Search Results
Model Number 40-3431
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.All information reasonably known as of 29 nov 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint: (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Event or Problem Description
Fda medwatch / fda user facility report# mw (b)(4) reported that on 29sep2022, the dobhoff was pulled, by the registered nurse (rn), and it was noted the tip on the dobhoff was not attached.The results from the kub (kidney, ureter, and bladder) noted the tip was broken; the medical doctor (md) was made aware.The kub was to be reordered; there was no reported injury.
 
Additional Manufacturer Narrative
The actual sample from the reported event was returned for evaluation.The provided sample confirmed the stylet and bolus offset (distal plug) component was missing; when viewed under magnification (30x) there was no visible bonding agent seen around the bolus assembly area.The reported event could be confirmed as reported; however, the root cause was not conclusively determined.The device history record for lot 30185293 was reviewed and the product was produced according to product specifications.All information reasonably known as of (b)(6) 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Event or Problem Description
It was reported, the tube had been inserted 9 days prior to the incident occurring; only liquid feeds and meds were provided.The tube was flushed following each feeding and prior to medication administration.The tip of the tube was left in the patient and per x-rays must have passed naturally.There was no history of this tube clogging prior to the incident.
 
Additional Manufacturer Narrative
The actual sample from the reported event was returned for evaluation, the stylet and bolus offset (distal plug) component were not returned with the provided sample.All information reasonably known as of 07 aug 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE W/STYLET W/ ENFIT CONNECTOR
Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key15885643
Report Number9611594-2022-00147
Device Sequence Number5099860
Product Code KNT
UDI-Device Identifier00350770460147
UDI-Public00350770460147
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K821906
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number40-3431
Device Lot Number30185293
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2022
Supplement Date Received by Manufacturer12/14/2022
07/26/2023
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date01/10/2023
08/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
Patient Weight65 KG
-
-