| Model Number |
850190 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that upon receipt of item, it was found that the protective layer over yellow foam was torn.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection of the returned device confirmed the defect.The cuff was torn.The complaint device was reviewed with the production operators, and a review of the production floor and the process was conducted to determine how the defect occurred and why it was not identified prior to packaging.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|