• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC INC FLEXI-SEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC INC FLEXI-SEAL Back to Search Results
Model Number 421630
Medical Device Problem Code Device Handling Problem (3265)
Health Effect - Clinical Code Tissue Breakdown (2681)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Common device name: tubes, gastrointestinal (and accessories).Procode: knt.Affiliation: (b)(6) medical center.Per additional information provided by the intake nurse, leakage was found to be around the tube, no product defect.Stool was reported as "watery, loose stool".Rectal exam's are not documented.Upon removal of fms device, poor rectal tone was noted at that time.Since no documentation recorded of rectal exam prior to use, it's unclear if the patient is a good candidate for use of the fms device.Per flexi-seal fecal management system (fms) ifu, a digital rectal exam is required prior to insertion.In addition, small amounts of moisture or seepage around the catheter is anticipated.No lot number is available.A detailed investigation or batch review cannot be conducted.Therefore, this evaluation will be closed.This issue will be monitored through the post market product monitoring review process.Based on the available information, this event is deemed to be a reportable serious injury.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: (b)(4).Manufacturing site: (b)(4).
 
Event or Problem Description
It was reported "that a patient using flexiseal protect fms developed a pressure injury to the perineum.It was also stated there "leakage is around the rectal tube" and they "are seeing an increase in hapis around rectal tubes".Staff is not consistent with use of casp mwf as their skin protectant.They do not irrigate daily.Per additional information received, casp was clarified to be "3m cavilon¿ advanced skin protectant ".A total of 14 patients were identified to have pressure injury to the perineum, however no information was provided for 13 of the patients although information has been requested.This emdr is covering patient #4 of 14 patients reported by this reporter.The remaining patients are covered in subsequent emdrs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXI-SEAL
Manufacturer (Section D)
CONVATEC INC
211 american avenue
greensboro NC
Manufacturer (Section G)
CONVATEC INC
211 american avenue
greensboro NC
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 
3365424679
MDR Report Key15885981
Report Number1049092-2022-00424
Device Sequence Number5122719
Product Code KNT
UDI-Device Identifier00768455143016
UDI-Public768455143016
Combination Product (Y/N)N
Initial Reporter StateIN
PMA/510(K) Number
K181206
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number421630
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/14/2022
Initial Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-