• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH EXTRACTOR ASNIS III SCREWS Ø6.5/8.0MM LARGE AO FITTING; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH EXTRACTOR ASNIS III SCREWS Ø6.5/8.0MM LARGE AO FITTING; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 702624
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the procedure the key broke when trying to remove it from the screw inside the patient.No additional information is available regarding if any debris was left behind in the patient.
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that during the procedure the key broke when trying to remove it from the screw inside the patient.No additional information is available regarding if any debris was left behind in the patient.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device, was returned and matches alleged failure.The device inspection revealed the following: the returned extractor screwdriver is broken from the tip, and the tip was not returned.From the microscopic image, it is clear that the fracture is brittle in nature and broken due to the application of torsional aspects.In the fractured area, we can see rust formation as started which is due to reprocessing after breakage.Furthermore, a hardness test according to rockwell on the returned item indicates the material being in accordance with the values in the material specification.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the investigation, the root cause was attributed to being user related.By application of a high torsional load, leading to a forced brittle fracture.If any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported that during the procedure the key broke when trying to remove it from the screw inside the patient.No additional information is available regarding if any debris was left behind in the patient.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device, was returned and matches alleged failure.The device inspection revealed the following: the returned extractor screwdriver is broken from the tip, and the tip was not returned.From the microscopic image, it is clear that the fracture is brittle in nature and broken due to the application of torsional aspects.In the fractured area, we can see rust formation as started which is due to reprocessing after breakage.Furthermore, a hardness test according to rockwell on the returned item indicates the material being in accordance with the values in the material specification.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the investigation, the root cause was attributed to being user related.By application of a high torsional load, leading to a forced brittle fracture.If any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTRACTOR ASNIS III SCREWS Ø6.5/8.0MM LARGE AO FITTING
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15886400
Report Number0008031020-2022-00661
Device Sequence Number5113652
Product Code LXH
UDI-Device Identifier04546540112873
UDI-Public04546540112873
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number702624
Device Catalogue Number702624
Device Lot Number15539Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/04/2022
Supplement Date Received by Manufacturer03/08/2023
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date04/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexMale
-
-