Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Halo (2227)
Date of Event 05/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that an intraocular lens (iol) was explanted from the patient's operative eye due to a mechanical complication-halos.The visual acuity (va) pre-operative: 20/50 and va post-operative: 20/20.The incision was enlarged and sutures and a vitrectomy were required.The patient is doing well post-operatively.No other information was provided.
 
Additional Manufacturer Narrative
Date of event: exact date unknown/not provided.Best estimate date is between (b)(6) 2022-(b)(6) 2022.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.(b)(4).An attempt has been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Corrected data: further information was provided and reported dr.(b)(6) was the explanting surgeon.It was initially reported as surgeon bing chiu.Additional info: through follow-up, the date of event was provided and the affected eye was the left eye.Another johnson & johnson lens, model zcb00 19.0 diopter was implanted as a replacement.The customer confirmed the product was returned to j&j for product evaluation.The following sections have been updated accordingly: section b3: date of event: (b)(6) 2022.Section e1: first/given name: (b)(6).Section e1: last name: (b)(6).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional info: section d9: device available for evaluation? yes.Section d9: returned to manufacturer on: jan 4, 2023.Section h3: device evaluated by manufacturer: yes.Device evaluation: the suspect lens was received an a bag with a foam swab.Visual inspection under magnification revealed that the lens was received cut in half.The lens was cleaned and inspected and lens damage and scratches were observed on the optics of the lens.The complaint issues were not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15886770
Report Number3012236936-2022-02943
Device Sequence Number17424020
Product Code MFK
UDI-Device Identifier05050474750654
UDI-Public(01)05050474750654(17)240912
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/12/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU190
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2022
Supplement Date Received by Manufacturer12/08/2022
01/23/2023
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date12/27/2022
02/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-