No device was returned and no photos could be provided confirming the alleged complaint.No lot code information was provided so a manufactural review could not be completed.Review of the reported event identified that the driver tip fractured during screw advancement indicating excessive off angled force and/or sclerotic bone as the likely root cause although no definitive cause could be determined.No additional investigation can be completed.Labeling review: "do not implant the instruments: complications to the patient may include, but are not limited to: breakage of the device, which could make necessary removal difficult or sometimes impossible, with possible consequences of late infection and migration.Breakage could cause injury to the patient." "pre-operative warnings - the methods of use for the instruments are to be determined by the user¿s experience and training in surgical procedures.Do not use these instruments for any action for which it was not intended such as hammering, prying, or lifting.These instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices".".The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury.The instruments should be inspected for damage prior to use.If they are bent or damaged in any way, they should not be used, regardless of navigated or manual procedures.In addition, periodically inspect the instruments for wear and tear, such as corrosion or discoloration.For instruments that are no longer functional, or exhibit excessive wear and tear, please return instruments to nuvasive." ".All sets should be carefully checked for completeness and all components should be carefully checked for signs of damage prior to use.Damaged packages or products should not be used and should be returned to nuvasive." ".Instruments should be protected during storage and from corrosive environments.All non-sterile parts should be cleaned and sterilized before use inspect all components for damage before use.Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient.Refer to cleaning and sterilization instructions below for all non-sterile parts." ".The nuvasive navigation instruments should not be bent or altered in any way.If damage is observed, the instrument should not be used and returned to nuvasive." ".Intra-operative warnings - the physician should take precautions against putting undue stress on the spinal area with instruments.Any surgical technique should be carefully followed.It is important that the surgeon exercise extreme caution when working in close proximity to vital organs, nerves, or vessels, and that the force applied to the instrumentation is not excessive, to prevent potential injury to the patient." "."the physical characteristics required for many instruments do not permit them to be manufactured from implantable materials.If any broken fragments of instruments remain in the body of a patient, they could cause allergic reactions or infections.If an instrument breaks in surgery and fragments go into the patient, these pieces should be removed prior to closure and should not be implanted." device not returned.
|