| Model Number |
145FS35 |
| Medical Device Problem Code |
Flaked (1246)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that guidewire coating seemed to be scratched off when going through the ureteroscope.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that guidewire coating seemed to be scratched off when going through the ureteroscope.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is unconfirmed as the problem could not be reproduced.One sample from lot gfgs1707 was returned in a polybag.The product was in its hoop.The unit was in good condition.No manufacturing issues or non-conformances were noted during review of the dhr that could have caused or contributed to the reported event.As the reported event is unconfirmed, a label/packaging review is not required.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the actual/suspected device was inspected.
|
| |
|