• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to olympus, the oes cystonephrofiberscope had a broken forceps plug cap.There was no report of user or patient harm associated with this event.
 
Additional Manufacturer Narrative
The subject device was returned to olympus for evaluation.During inspection and testing, the allegation was confirmed.The forceps stopper cap came off.In addition, water tightness was not maintained due to perforation of the forceps channel.There were stains on the image guide bundle.Additional information has been requested regarding this event.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if additional information becomes available.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and additional information received.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be determined.It is likely the forceps port was damaged due to stress, handling, or other factors.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
Additional information was provided.The issue occurred during reprocessing after a therapeutic stent exchange procedure.There were no procedural delays and the procedure was completed using the same device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Common Device Name
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15887025
Report Number3002808148-2022-04709
Device Sequence Number5123231
Product Code FAJ
UDI-Device Identifier04953170292521
UDI-Public04953170292521
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K032092
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCYF-5A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2022
Supplement Date Received by Manufacturer12/05/2022
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date12/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AMM215WTD (8FMZ0011); CH-S190-08-LB (7806673); CLV-S190 (7826031); OTV-S190 (7805865)
Patient SexUnknown
-
-