DEPUY ORTHOPAEDICS INC US SUMMIT DUOFIX TAP SZ5 HI OFF; HIP FEMORAL STEM Back to Search Results
Model Number 1570-12-110
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Synovitis (2094); Distress (2329); Metal Related Pathology (4530); Unspecified Tissue Injury (4559); Physical Asymmetry (4573)
Date of Event 05/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.The device catalog number is unknown; therefore, udi # is unavailable.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Pinnacle mom litigation record received.Litigation alleges heavy metal poisoning from toxic heavy metal resulting to injuries, hip pain, metallosis, metal wear and limited adl.Plaintiff also suffered emotional trauma and distress.Plaintiff suffered subsequent dislocation post revision left hip.Doi: (b)(6) 2008; dor: (b)(6) 2022; left hip.
 
Event or Problem Description
Medical records received.After review of the medical records, the patient was revised to address failed left tha due to metal-on-metal severe metallosis and elevated metal ions.Operative note reported scarred tissue and synovial fluid.There was extensive dark gray reactive synovitis, pseudotumor tissues, and a mild amount of osteolysis around the femur.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H10 additional narrative: added: a2 (age, dob), a3, b5, b7, d4 (lot, catalog, expiry date, udi), g4 (pma/ 510(k)), h4, h6 health effect - clinical code corrected: a1, d1, d2a, d10 (concomitant).
 
Event or Problem Description
Medical records received.In addition to what was previously reported in the medical records patient had a disability and limited activity.The operative note reported scarred tissue flap of the posterior greater trochanter and found a moderately compromised dehisced posterior soft tissue envelope.The left leg is longer than the right by 7 mm and increased complexity, effort, and time due to the severity of the patient's pathology.Lab results for metal ions are above 7 ppb.Bilateral hip implant, however, the right hip had a competitor implant.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary the device associated with this report was not returned to depuy synthes for evaluation.Review of the x-ray evidence found nothing indicative of a device nonconformance.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT DUOFIX TAP SZ5 HI OFF
Common Device Name
HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15887528
Report Number1818910-2022-23991
Device Sequence Number5123249
Product Code LPH
UDI-Device Identifier10603295060161
UDI-Public10603295060161
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K193398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2018
Device Model Number1570-12-110
Device Catalogue Number157012110
Device Lot NumberB95CVA000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2022
Supplement Date Received by Manufacturer03/09/2023
03/28/2023
08/07/2023
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date03/09/2023
04/05/2023
08/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/10/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICULEZE M HEAD 36MM +1.5; PINN CAN BONE SCREW 6.5MMX40MM; PINNACLE MTL INS NEUT36IDX52OD; PINNACLE SECTOR II CUP 52MM; UNK HIP ACETABULAR CUP PINNACLE; UNK HIP ACETABULAR LINER METAL PINNACLE; UNK HIP FEMORAL HEAD METAL
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexFemale
Patient Weight77 KG
Patient RaceWhite
-
-