Model Number 36725 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/08/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the procedure was not completed due to this event.A rf3000 radiofrequency generator was selected for use.During the procedure, the generator did not work after turning on.The procedure was not completed due to the event.There were no patient complications or additional intervention, and the patient status is stable.
|
|
Event Description
|
It was reported that the procedure was not completed due to this event.A rf3000 radiofrequency generator was selected for use.During the procedure, the generator did not work after turning on.The procedure was not completed due to the event.There were no patient complications or additional intervention, and the patient status is stable.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: the device was returned and serviced.The device failed the post test with a h07 fault.The rf board was replaced with a good board and the device passed the post testing.This unit is to be scrapped.
|
|