• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE VR ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE VR ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD1311-36Q
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Heart Failure/Congestive Heart Failure (4446)
Date of Event 11/01/2022
Type of Reportable Event Death
Event or Problem Description
Related manufacturer reference number: 2017865-2022-47193.It was reported that the patient expired on (b)(6) 2022.There is no known allegation from a health professional that suggests the death was related to the device.It was reported that the cause of death was congestive heart failure.
 
Additional Manufacturer Narrative
Interrogation of the device revealed it was above elective replacement indicator (eri) when received.The device¿s sensing, pacing, lead impedance, high voltage charging, and high voltage shock impedance were tested, and the device¿s function was normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE VR ICD
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15888074
Report Number2017865-2022-47192
Device Sequence Number5044146
Product Code NVZ
UDI-Device Identifier05414734506113
UDI-Public05414734506113
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2014
Device Model NumberCD1311-36Q
Device Catalogue NumberCD1311-36Q
Device Lot Number3810065
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2022
Supplement Date Received by Manufacturer01/06/2023
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date01/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/12/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age82 YR
Patient SexFemale
-
-