MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ULTRAFLEX IAB: 7.5FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ULTRAFLEX IAB: 7.5FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN039788
Medical Device Problem Codes Connection Problem (2900); Device Misassembled During Manufacturing /Shipping (2912)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tubing connector of the helium adaptor was found upside down.Further information states that "[the clinician] had already inserted the catheter into the patient, then the connection between our adaptor and port of the maquet iabp was found after they wanted to plug in and started pumping".No report of harm or injury to the patient.
 
Additional Manufacturer Narrative
(b)(4).Other remarks: n/a.Corrected data: n/a.
 
Additional Manufacturer Narrative
Qn# (b)(4).The reported complaint that the "tubing connector of the helium adaptor was found upside down" is confirmed based on the customer photos provided with the complaint report.The data-scope connector was noted in a backwards orientation within the supplied driveline tubing.The product was not returned for investigation.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the backwards driveline connector.The root cause is manufacturing related.A non-conformance has been initiated to further investigate the issue.
 
Event or Problem Description
It was reported that the tubing connector of the helium adaptor was found upside down.Further information states that "[the clinician] had already inserted the catheter into the patient, then the connection between our adaptor and port of the maquet iabp was found after they wanted to plug in and started pumping".No report of harm or injury to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX IAB: 7.5FR 40CC
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
MDR Report Key15889814
Report Number3010532612-2022-00536
Device Sequence Number5219090
Product Code DSP
UDI-Device Identifier10801902145731
UDI-Public10801902145731
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K000729
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2024
Device Model NumberIPN039788
Device Catalogue NumberIAB-06840-U
Device Lot Number18F22C0038
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2022
Supplement Date Received by Manufacturer12/22/2022
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-