| Model Number |
468-450 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/31/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right deflation and "textured to smooth due to product concern".Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported, right deflation.And "textured to smooth, due to product concern".Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: anxiety-product/procedure: unable to confirm, as it is a medical event.Not related to the device.Deflation: observed, a sharp assessed, as unidentified (tear) opening (shell thickness was within specification).Also observed, a delamination partial in the plug strap disk shell assessed, as to cohesive failure.And broken sharp in the plug strap assessed, as to unidentified (tear) opening.Additional observations: crease flat observed, in the surface.Yellow particles observed, in the surface of the device.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|