AOMORI OLYMPUS CO., LTD. SONICBEAT 5 MM, 20 CM, FRONT-ACTUATED GRIP; INSTRUMENT Back to Search Results
Model Number SB-0520FC
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is returned, and an evaluation completed for it.The user¿s complaint was confirmed.Upon inspection of the device, it was observed that the tissue pad was worn, partly peeled and torn.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
 
Event or Problem Description
As reported for this event by the customer, during a therapeutic laparotomy procedure, the device tissue pad peeled.The procedure was completed with a new similar device.There is no harm or adverse impact to the patient.The intelligent tissue monitoring (itm) setting was off during the procedure.
 
Additional Manufacturer Narrative
Additional information has been received from the customer.This supplemental report is being submitted to provide this information as well as additional information based on the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although a definitive root cause cannot be identified, the following step-by-step scenario likely that the ultrasonic output was possibly activated when the grasping section was not grasping anything (this includes after tissue resection).This might have caused the tissue pad to wear out and partially peeled off.The following information is included in the instructions for use (ifu): ¿do not activate output while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ ¿when cutting or vessel sealing is performed, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ olympus will continue to monitor the performance of this device.
 
Event or Problem Description
The customer provided additional information on 13dec2022: the event occurred towards the end of the procedure.The functional mode and settings at the time of the event are not known.No part of the tissue pad came off.There was minimal delay to the completion of the procedure, just enough to replace the device with a new one.The patient did not need to be given additional anesthesia.It is not known if there was any error message or alarm received for the device at the time of the event.The device was inspected prior to use with no anomaly noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONICBEAT 5 MM, 20 CM, FRONT-ACTUATED GRIP
Common Device Name
INSTRUMENT
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15890313
Report Number9614641-2022-00681
Device Sequence Number5218594
Product Code LFL
UDI-Device Identifier04953170370601
UDI-Public04953170370601
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K211838
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSB-0520FC
Device Lot Number1YK 01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date12/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TD-SB400ESG-400USG-400
Patient SexUnknown
-
-