| Model Number |
SB-0520FC |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is returned, and an evaluation completed for it.The user¿s complaint was confirmed.Upon inspection of the device, it was observed that the tissue pad was worn, partly peeled and torn.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
|
| |
|
Event or Problem Description
|
|
As reported for this event by the customer, during a therapeutic laparotomy procedure, the device tissue pad peeled.The procedure was completed with a new similar device.There is no harm or adverse impact to the patient.The intelligent tissue monitoring (itm) setting was off during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information has been received from the customer.This supplemental report is being submitted to provide this information as well as additional information based on the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although a definitive root cause cannot be identified, the following step-by-step scenario likely that the ultrasonic output was possibly activated when the grasping section was not grasping anything (this includes after tissue resection).This might have caused the tissue pad to wear out and partially peeled off.The following information is included in the instructions for use (ifu): ¿do not activate output while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ ¿when cutting or vessel sealing is performed, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ olympus will continue to monitor the performance of this device.
|
| |
|
Event or Problem Description
|
|
The customer provided additional information on 13dec2022: the event occurred towards the end of the procedure.The functional mode and settings at the time of the event are not known.No part of the tissue pad came off.There was minimal delay to the completion of the procedure, just enough to replace the device with a new one.The patient did not need to be given additional anesthesia.It is not known if there was any error message or alarm received for the device at the time of the event.The device was inspected prior to use with no anomaly noted.
|
| |
|