• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL EZ IO 45MM NEEDLE SET + STABILIZER; NEEDLE, HYPODERMIC, SINGLE LU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL EZ IO 45MM NEEDLE SET + STABILIZER; NEEDLE, HYPODERMIC, SINGLE LU Back to Search Results
Model Number IPN917352
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event or Problem Description
Medwatch (b)(4).Patient arrived as trauma activation on (b)(6) 2022 from the scene of a motorcycle crash.Patient was transported by lifenet medevac who placed intraosseous catheter in right humerus at 1540 pta in ed for access.An a-line and central line were placed in ed prior to or.No documentation of intraosseous catheter removal.On (b)(6) 2022 patient complained of pain in right shoulder with occupational therapy.X-rays ordered by ortho resident with note of retained foreign body, io needle.No documentation found of when io was removed.Typically, these are removed once central access obtained.Patient sent to or on (b)(6) 2022 to remove retained catheter needle, with no further complications.
 
Event or Problem Description
Medwatch (b)(4).Patient arrived as trauma activation on 10/2 from the scene of a motorcycle crash.Patient was transported by (b)(6) who placed intraosseous catheter in right humerus at 1540 pta in ed for access.An a-line and central line were placed in ed prior to operating room.No documentation of intraosseous catheter removal.On 10/12 patient complained of pain in right shoulder with occupational therapy.X-rays ordered by ortho resident with note of retained foreign body, io needle.No documentation found of when io was removed.Typically, these are removed once central access obtained.Patient sent to operating room on 10/15 to remove retained catheter needle, with no further complications.
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZ IO 45MM NEEDLE SET + STABILIZER
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LU
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key15890626
Report Number3011137372-2022-00215
Device Sequence Number5209994
Product Code FMI
UDI-Device Identifier10801902170641
UDI-Public10801902170641
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
PMA/510(K) Number
K180395
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2025
Device Model NumberIPN917352
Device Catalogue Number9079P-VC-005
Device Lot Number55675249
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2022
Supplement Date Received by Manufacturer12/13/2022
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED; NONE REPORTED
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age38 YR
Patient SexMale
-
-