• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD Q-SYTE 15CM SMALL BORE TRI-EXT SETS; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD Q-SYTE 15CM SMALL BORE TRI-EXT SETS; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 385158
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that 3 bd q-syte 15cm small bore tri-ext sets were not able to prime.The following information was provided by the initial reporter: when the user is trying to prime the q syte tri extension 385158, not able to prime two lumens due to resistance in septum and also not able to connect bd posiflush lue lok to q syte.One lumen we can access.
 
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that 3 bd q-syte 15cm small bore tri-ext sets were not able to prime.The following information was provided by the initial reporter: when the user is trying to prime the q syte tri extension 385158, not able to prime two lumens due to resistance in septum and also not able to connect bd posiflush lue lok to q syte.One lumen we can access.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 12-dec-2022.H6: investigation summary our quality engineer inspected the 6 samples submitted for evaluation.The reported issues of flow issues - fluid blockage and connection issues were confirmed upon inspection of the samples.Analysis of the sample showed that the top and bottom membrane disks of the q-syte did not have the proper slit opening.Bd determined that the cause of the failure was associated to the manufacturing process.During the manufacturing process the disks of the q-sytes have a slit cut into them via a blade.If the blade is dull, broken, or missing in the manufacturing machinery than the slit will not be properly cut into the disk resulting in flow issues and connection issues.The device history records (dhr) review was performed for the lot number material identified in this complaint.According to the documented records, the product was manufactured according to the approved manufacturing procedures, specifications, and then released by quality assurance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Q-SYTE 15CM SMALL BORE TRI-EXT SETS
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15890738
Report Number9610847-2022-00447
Device Sequence Number20505654
Product Code FPA
UDI-Device Identifier00382903851584
UDI-Public00382903851584
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K142527
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/31/2025
Device Catalogue Number385158
Device Lot Number2033442
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2022
Supplement Date Received by Manufacturer01/20/2023
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date02/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-