| Model Number |
90186 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not available to the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that during an endovascular procedure there was resistance when the subject stent retriever was advanced through the microcatheter and when the physician tried to return the pusherwire, the subject stent retriever broke.There was a surgical delay of unknown duration due to the event.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.No other information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
There are controls in the manufacturing process to ensure the product met specifications upon release.During visual/microscopic inspection, the retriever core wire was seen to be broken approx.168.5cm from the proximal end.The retriever coils were seen to be damaged.The retriever core wire was seen to be kinked/bend.The retriever shaped section was intact.The insertion tool was not returned.During functional inspection, the retriever coil was stretched during the analyses to reveal the core wire breaking point.Functional inspection for 'retriever difficult/unable to go through catheter shaft' could not be carried out due to the damage noted to the device.During sem (scanning electron microscopy) rd (research & development team) analysis, saw handling damage around the edge, especially in proximal end of the fracture.The fracture belongs to ductile overload failure under bending load condition.Cracks initiated at two locations at compression side.Saw evidence of hydrogen embrittlement at the edge near the initiation locations.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported events retriever fracture/broken during use and retriever difficult/unable to go through catheter shaft could not be confirmed; however, the analysis results are consistent with the reported event.Additional information was received indicating that the device was prepared for use as per the directions for use, no damage noted to the packaging prior to opening the packaging, device was confirmed to be in good condition during preparation/prior to use on the patient, and continuous flush set up and maintained throughout the clinical procedure.In addition, the trevo xp retriever was used with a non-stryker microcatheter and experienced big friction in microcatheter, which may have contributed to the reported event.An assignable cause of procedural factors will be assigned to the as reported 'retriever fractured/broken during use' and 'retriever difficult/unable to go through catheter shaft', and the as analyzed code 'retriever core wire broken during use', 'retriever coils damaged' and 'retriever core wire kinked' as the issue is associated with a product that meets stryker design and manufacture specifications and was used in according with the dfu but due to procedural and/or anatomical factors during use, the product performance was limited.
|
| |
|
Event or Problem Description
|
|
It was reported that during an endovascular procedure there was resistance when the subject stent retriever was advanced through the microcatheter and when the physician tried to return the pusherwire, the subject stent retriever broke.There was a surgical delay of unknown duration due to the event.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.No other information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H4 manufacturing date ¿ added.D4 expiration date - added.There are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, visual testing as well as functional testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information was received indicating that the device was prepared for use as per the directions for use, no damage noted to the packaging prior to opening the packaging, device was confirmed to be in good condition during preparation/prior to use on the patient, and continuous flush set up and maintained throughout the clinical procedure.In addition, the subject stent retriever was used with a non-stryker microcatheter and experienced big friction in mc, which may have contributed to the reported event.An assignable cause of procedural factors will be assigned to the as reported 'retriever fractured/broken during use' and 'retriever difficult/unable to go through catheter shaft', as this issue appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural and/or anatomical factors during use.
|
| |
|
Event or Problem Description
|
|
It was reported that during an endovascular procedure there was resistance when the subject stent retriever was advanced through the microcatheter and when the physician tried to return the pusherwire, the subject stent retriever broke.There was a surgical delay of unknown duration due to the event.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.No other information is available.
|
| |
|