• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD RESPIRATORY HUMIDIFIER; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD RESPIRATORY HUMIDIFIER; BTT Back to Search Results
Model Number MR850
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The complaint mr850 respiratory humidifier is currently en-route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of investigation.
 
Event or Problem Description
A distributor reported on behalf of a healthcare facility in thailand via a fisher & paykel healthcare (f&p) field representative that the audible alarm of a mr850 respiratory humidifier was not functioning properly.There was no patient involvement.
 
Additional Manufacturer Narrative
(b)(4).Method: the pcb of the complaint mr850 respiratory humidifier was returned to fisher & paykel healthcare in new zealand for investigation, where it was electrically evaluated.Results: performance testing revealed that the audible alarm of the mr850 was functioning as intended.Conclusion: we are unable to determine the cause of the reported event.Our mr850 product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 heater base.In addition, the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 is equipped with visual alarm indicators in addition to the audible alarm.
 
Event or Problem Description
A distributor reported on behalf of a healthcare facility in thailand via a fisher & paykel healthcare (f&p) field representative that the audible alarm of a mr850 respiratory humidifier was not functioning properly.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRATORY HUMIDIFIER
Common Device Name
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534000
MDR Report Key15891580
Report Number9611451-2022-01106
Device Sequence Number20138340
Product Code BTT
UDI-Device Identifier09420012407207
UDI-Public(01)09420012407207(10)2101951605(11)211213
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR850
Device Catalogue NumberMR850
Device Lot Number2101951605
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2022
Supplement Date Received by Manufacturer07/13/2023
Initial Report FDA Received Date11/30/2022
Supplement Report FDA Received Date07/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-