| Model Number |
ZA9003 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable, as lens was not implanted.If explanted, give date: not applicable, as lens was not implanted, hence not explanted.The device is not returned; therefore, a failure analysis of the complaint device cannot be completed.A review of the device record and complaint trending for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
Report of intraocular lens (iol) when inserting partially inserted, haptic was bent, used back-up iol to complete.There was no patient injury, patient doing fine.No other information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9: device available for evaluation? yes.Returned to manufacturer on: 12/28/2022.Section h3: device evaluated manufacturer? yes.Device evaluation: visual inspection under magnification revealed that the complaint lens was received with one of the haptics damaged.The lens was cleaned and, no further issues were identified, the complaint issue " dc-haptic damaged" was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Conclusion: as a result of the investigation, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|