| Model Number |
CD3365-40C |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient presented remotely via merlin.Net.Upon review, the implantable cardioverter defibrillator exhibited electrogram anomaly; a morphology template electrogram was recorded on a device where the morphology template discriminator and consequently the episode trigger was turned off.No intervention was performed.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of a recorded morphology template egm when the trigger was turned off was confirmed.Based on the information provided, it was determined that an automatic template maintenance was initiated; thus, an egm was triggered.A potential template can be created even if the morphology discriminator is disabled.The device was performing as designed.
|
| |
|