• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC PIVET GUIDE EMBRYO TRANSFER SET; MQF CATHETER, ASSISTED REPRODUCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC PIVET GUIDE EMBRYO TRANSFER SET; MQF CATHETER, ASSISTED REPRODUCTION Back to Search Results
Model Number G30457
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) number k173103.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
As reported, during an in vitro fertilization (ivf) procedure, a pivet guide embryo transfer set catheter generated bubbles (b)(4) in the inner catheter and "uneven or black spots" were noticed prior to aspiration under a microscope in the front one-third section of the catheter.A bd syringe and vitrolife blastocyst culture medium were used.There were no adverse effects reported and the procedure was completed by using another new device of the same type.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: this report is being sent to indicate the complaint event is not reportable under fda 21 cfr part 803.As such, this event no longer meets the set criteria for a reportable event; no further reports regarding this event will be submitted.Investigation has determined the alleged malfunction corresponds with surface defect instead of foreign matter, which our initial report was based on.Event is not reportable under fda 21 cfr part 803 as it does not meet the criteria for a death or serious injury and there is no precedence of this malfunction leading to an adverse event.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIVET GUIDE EMBRYO TRANSFER SET
Common Device Name
MQF CATHETER, ASSISTED REPRODUCTION
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15893033
Report Number1820334-2022-01791
Device Sequence Number6690026
Product Code MQF
UDI-Device Identifier00827002304574
UDI-Public(01)00827002304574(17)250713(10)14839531
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/13/2025
Device Model NumberG30457
Device Catalogue NumberK-PETS-5000
Device Lot Number14839531
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/14/2022
Supplement Date Received by Manufacturer02/08/2023
Initial Report FDA Received Date12/01/2022
Supplement Report FDA Received Date03/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-