| Model Number |
ALLURA XPER FD10 OR TABLE |
| Medical Device Problem Code |
Radiation Output Failure (4027)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It has been reported to philips that the footswitch and fluoroscopy was not functioning.The issue was found during clinical use.No harm has been reported to philips.Philips has started an investigation of this complaint.
|
| |
|
Additional Manufacturer Narrative
|
|
Philips has investigated this complaint.The philips field service engineer (fse) inspected the system onsite and confirmed that the fluoroscopy was not functioning due to the damaged footswitch cable.Fse replaced the wired footswitch and after replacement, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
|
| |
|