STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 36MM/+10; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 6260-9-336
Medical Device Problem Codes Degraded (1153); Material Erosion (1214)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
 
Event or Problem Description
As reported: "right hip revision due to gross trunnion failure." rep provided prior usage and an explant picture showing trunnion wear.Rep confirmed that while the femoral head was easily removed, it wasn't dissociated.Rep also confirmed that no further information will be released by the hospital or surgeon.
 
Additional Manufacturer Narrative
Reported event: an event regarding wear involving a metal head was reported.The event was confirmed via evaluation of the provided photograph of the stem.Method & results: -product evaluation and results: the reported device was not returned however a photograph of the stem was provided for review.Visual inspection of the provided photograph indicated that the stem is severely worn consistent with loss of taper lock.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient's hip was revised due to trunnion wear leading to loss of taper lock with the head.The subject device has been identified to be within scope of nc and capa.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of nc and capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analyses identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.No further investigation is required.
 
Event or Problem Description
As reported: "right hip revision due to gross trunnion failure." rep provided prior usage and an explant picture showing trunnion wear.Rep confirmed that while the femoral head was easily removed, it wasn't dissociated.Rep also confirmed that no further information will be released by the hospital or surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 36MM/+10
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15893721
Report Number0002249697-2022-01750
Device Sequence Number5210244
Product Code JDI
UDI-Device Identifier07613327032321
UDI-Public07613327032321
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K022077
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2004
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2009
Device Model Number6260-9-336
Device Catalogue Number6260-9-336
Device Lot Number13695301
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer01/21/2023
Initial Report FDA Received Date12/01/2022
Supplement Report FDA Received Date02/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/22/2004
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2249697-05/07/2018-003R
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age87 YR
Patient SexMale
Patient Weight86 KG
-
-