• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP Back to Search Results
Model Number e1801-44
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product involved in the reported issue was discarded at the hospital and therefore not available for evaluation.The root cause of the reported issue could not be determined.We have reviewed the device history record for the involved lots.No issues were found during manufacturing or packaging that would cause or contribute to the reported event.Additionally, we have not received any reports of similar issues occurring for the involved lots.Please note that this is report 1 of 4 related to this incident.Reports 1220948-2022-00125, 1220948-2022-00126 and 1220948-2022-00127 were also filed in relation to the incident.
 
Event or Problem Description
A customer reported various issues have occurred while using the syntel silicone embolectomy catheter - regular tip devices.It was reported that the user had the balloon of the catheter rupture during use and they had to retrieve the broken piece with a snare.Additionally, the site reported three additional incidences of failures with this product.Two additional devices also had balloons rupture during use and one device had the balloon fail to deflate during use.No injuries were reported for these events.Please note that this is report 1 of 4 related to this incident.Reports 1220948-2022-00125, 1220948-2022-00126 and 1220948-2022-00127 were also filed in relation to the incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP
Common Device Name
EMBOLECTOMY CATHETER
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer Contact
jeffrey oddy
63 second ave
burlington, MA 01803
7814251682
MDR Report Key15894158
Report Number1220948-2022-00124
Device Sequence Number15151688
Product Code DXE
UDI-Device Identifier00840663101030
UDI-Public(01)00840663101030
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K992368
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Numbere1801-44
Device Catalogue NumberE1801-44
Device Lot NumberNSE2440 AND NSE2513
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2022
Initial Report FDA Received Date12/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-