| Model Number |
CRPLUS |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.Patient fields in which information is not provided were intentionally left blank.Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device failed to power on.In this state, the device may not be able to deliver defibrillation therapy if needed.This issue is patient related.However; there was no adverse event reported.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device failed to power on.In this state, the device may not be able to deliver defibrillation therapy if needed.This issue is patient related; however, there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customers device and was able to duplicate the reported issue.Root cause of the reported failure has determined to be the detached lid latch component.The returned device was archived by stryker.The customer received a replacement device.
|
| |
|