MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH URETEROSCOPE, 7,5 FR. X 330 MM, 7°, 2 CHANNELS, STRAIGHT OCULAR ANGLE; SEMI-RIGID URETEROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH URETEROSCOPE, 7,5 FR. X 330 MM, 7°, 2 CHANNELS, STRAIGHT OCULAR ANGLE; SEMI-RIGID URETEROSCOPE Back to Search Results
Model Number WA02946A
Medical Device Problem Codes Break (1069); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
Olympus was informed via service request by the sterile processing manager at the user facility that during reprocessing, the customer dual channel semi-rigid ureteroscope reportedly was found to have ¿broken rod lens and the shaft is bent¿.No death, injury, or harm to the patient or other was reported to olympus.
 
Additional Manufacturer Narrative
The subject device was returned to the olympus repair center.The repair inspection confirmed the customer¿ reported issue, the distal objective cover glass lens is cracked and the outer tube shaft is bent.The image is blurry.No moisture was found.The investigation is ongoing.However, if additional information becomes available, this report will be supplemented accordingly.In general, the customer is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 3 years since the subject device was manufactured.Based on the results of the investigation, it is likely the reported event occurred due to excessive use.However, a definitive root cause could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETEROSCOPE, 7,5 FR. X 330 MM, 7°, 2 CHANNELS, STRAIGHT OCULAR ANGLE
Common Device Name
SEMI-RIGID URETEROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15895835
Report Number9610773-2022-00571
Device Sequence Number14754897
Product Code FGB
UDI-Device Identifier04042761029940
UDI-Public04042761029940
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K951855
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA02946A
Device Catalogue NumberWA02946A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2022
Supplement Date Received by Manufacturer03/01/2023
Initial Report FDA Received Date12/01/2022
Supplement Report FDA Received Date03/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/20/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-