| Model Number |
MT22949 |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that signal loss over one hour occurred.No product or data was provided for evaluation.The allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that signal loss over one hour occurred.The product was evaluated.An exterior visual inspection was performed and passed.Charge and boot testing was performed and passed.Functional test was performed and passed.Pairing test was performed and passed.The receiver log was downloaded and reviewed finding no errors related to the complaint.The allegation could not be determined.A probable cause could not be determined as the allegation was not found.No injury or medical intervention was reported.
|
| |
|