Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: date of event: only the event year is known.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Occupation: reporter is a j&j employee.The photo was returned to depuy synthese for evaluation.The depuy synthese team conducted a visual inspection of the returned photo.Visual analysis of the photo revealed that the speedshift comp impl kit 15x20x20 ofst 8 had the staple unattached inside the sealed package, this condition is most likely due to breakage of the clamps, staple and fragment are visible in the evidence provided.The observed condition of the speedshift comp impl kit 15x20x20 ofst 8 was consistent with a design related issue.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthese quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the speedshift comp impl kit 15x20x20 ofst 8.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|