• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMPLANT DIRECT SYBRON MANUFACTURING LLC SIMPLYLEGACY DRIVER; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMPLANT DIRECT SYBRON MANUFACTURING LLC SIMPLYLEGACY DRIVER; DENTAL IMPLANT Back to Search Results
Catalog Number HT3.0LG
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/03/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Per complaint (b)(4), during clinical procedure, components could not be separated.
 
Additional Manufacturer Narrative
Patient age is unknown.Implant date and explant date are not applicable since the product was never placed and not removed.Lot information is unknown.Device evaluation results are not available.If the analysis is complete, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLYLEGACY DRIVER
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
IMPLANT DIRECT SYBRON MANUFACTURING LLC
3050 east hillcrest drive
thousand oaks CA 91362
Manufacturer Contact
marcos gutierrez
3050 east hillcrest drive
thousand oaks, CA 91362
8184443315
MDR Report Key15896394
Report Number3001617766-2022-05901
Device Sequence Number5258135
Product Code NDP
UDI-Device Identifier10841307119890
UDI-Public10841307119890
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Dentist
Remedial Action Inspection
Type of Report Initial
Report Date (Section B) 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberHT3.0LG
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/03/2022
Initial Report FDA Received Date12/01/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight68 KG
-
-