| Model Number |
0210 |
| Medical Device Problem Code |
Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Reportable as if it were to recur it would require medical intervention.
|
| |
|
Event or Problem Description
|
|
Will not nebulize.
|
| |
|
Additional Manufacturer Narrative
|
|
Reportable as if it were to recur it would require medical intervention.The root cause cannot be definitely determined with the information given.Failure to nebulize could be from defective molding, or if the bottom of the nebulization cup is not screwed in all the way.Reviewed the dhrs and no causes for the reported defect were found.Reviewed the complaint history and there was no trending.Performed risk analysis and the severity rating of the complaint is 6/10.Sent resolution to the customer.
|
| |
|
Event or Problem Description
|
|
Will not nebulize.
|
| |
|