According to the reporter, the device was placed on the patient¿s feet overnight.Blisters and red spots were noticed on the patient¿s foot the following morning where the device was placed.The device was rotated onto the other foot the next day.However, blisters and spots were noticed again on the other foot where the device was placed.It was reported that the blisters looked like they could have been caused by heat.The skin condition was apparent immediately after removal.A self-adhesive black medical tape (to help disperse the heat) was used over the top of the device to keep it in place.The device was not checked during the 8-12 hours when the device was placed on the patient.The skin of the patient was not cleaned before the application of the device.The patient was placed on supine position and had a little pressure on the foot from swaddle.The patient has no allergies.The blisters were not getting worse; one of the blisters had popped and was resolving.
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|