• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MASTERS SERIES HEART VALSALVA AORTIC VALVES GRAFT; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MASTERS SERIES HEART VALSALVA AORTIC VALVES GRAFT; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 27VAVGJ-515
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cardiac Arrest (1762); Tachycardia (2095); Ventricular Fibrillation (2130)
Date of Event 11/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
Clinical study patient id: (b)(4).It was reported that on (b)(6) 2022, a 27mm sjm masters series valsalva aortic valved graft was successfully implanted in a patient.It was noted that at an unknown point after implant the patient experienced ventricular fibrillation that was preceded by ventricular tachycardia.The patient went into cardiac arrest and had to be shocked with 300 joules for defibrillation.The patient was administered magnesium sulfate, potassium chloride, lignocaine, and given a cordaron infusion.It was noted that after a few minutes the patient again experienced ventricular fibrillation.The patient was again defibrillated, but this time with only 200 joules.The patient's cordaron infusion was increased and administration of lignocaine was changed to intravenous.The patient's condition improved after treatment and there was no longer any ventricular fibrillation detected.The event resulted in the patient becoming hospitalized to monitor their rhythm.The patient's ventricular fibrillation resolved without sequelae.The patient was stable and discharged at the time of report.
 
Additional Manufacturer Narrative
An event of ventricular fibrillation, ventricular tachycardia and cardiac arrest was reported.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Information from the field indicated that the field believed the event was related to procedure or patient comorbidity.There was no allegation of malfunction against the abbott device.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Event or Problem Description
Clinical study patient id:(b)(6).Subsequent to the previously filed report, additional information was received that it is thought that the event is related to procedure or patient comorbidity, no allegation of malfunction against abbott device.The patient remained hemodynamically stable throughout the procedure and there was no clinically significant delay in procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASTERS SERIES HEART VALSALVA AORTIC VALVES GRAFT
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15897729
Report Number2135147-2022-02303
Device Sequence Number6793894
Product Code LWQ
UDI-Device Identifier05414734009515
UDI-Public05414734009515
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/01/2024
Device Model Number27VAVGJ-515
Device Catalogue Number27VAVGJ-515
Device Lot Number8479484
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer03/09/2023
Initial Report FDA Received Date12/01/2022
Supplement Report FDA Received Date03/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age49 YR
Patient SexMale
Patient Weight69 KG
Patient RaceWhite
-
-