| Model Number |
4.0PDC |
| Medical Device Problem Codes |
Use of Device Problem (1670); Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
Unintended Extubation (4564)
|
| Date of Event |
11/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, at home during use, the product was tested prior to use.The patient keeps having issues with the pilot balloon.It has been breaking off and subsequently cuff not holding air.The event required a replacement of the tube.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the device was missing the inflation line, and the inflation line was broken at the connection to the flange-connector.Functionally, a syringe needle was connected to the part of inflation line to test the cuff and air was applied and it was observed the cuff held the air and no leaks were observed when submerged into a container filled with water.It was reported that there was a leak in the cuff without an accompanying report on whether the customer attempted inflation, and there was an issue with the pilot line or balloon.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: avoid pulling or manipulation of the inflation line, as it is designed to conduct and hold air as part of the cuff inflation system.It is recommended that the inflation line be maintained in a position allowing for patient mobility without placing tension on the line-to-flange junction.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|