| Model Number |
1012455-08 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaints appear to be related to operational context.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a 85% stenosed lesion in the left anterior descending (lad) artery with heavy calcification and moderate tortuosity.The 5.00x8mm nc trek balloon dilatation catheter (bdc) was used for post dilatation; however, a kink was observed.Therefore the bdc was removed from the patient and difficulty was noted with the guide catheter.The bdc and the guide catheter had to be removed as a single unit.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
|
| |
|