| Model Number |
N/A |
| Medical Device Problem Codes |
Leak/Splash (1354); Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the surgery the air leakage from the connection of the tube and the cuff was found on this complaint product.There was a 0-15 minute delay.No adverse events were reported as a result of this malfunction.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This event is recorded with zimmer biomet under (b)(4).Once the investigation has been completed, a follow-up mdr will be submitted.(b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
This event has been recorded by zimmer biomet under: (b)(4).This medwatch is being filed to relay additional information.When connected to a known working ats in the lab, the ats had to keep pumping air to keep the cuff inflated, confirming a leak.A bubble test determined the leak was located where the pvc tubing connects to the right-angle connectors of the cuff.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further information is available at this time.
|
| |
|