Model Number A2030-200
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other incidents and/or complaints reported from this lot.Based on the information provided, the balloon rupture appears to be due to circumstances of the procedure.It is likely that the rupture occurred due to interaction with lesion and/or associated devices resulting in rupture during inflation.There were no leaks noted during preparation indicating that the damage was not pre-existing.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The additional armada 14 device referenced is filed under separate medwatch report numbers.
 
Event or Problem Description
It was reported that the procedure was to treat a left posterior tibial artery.A 2x200mm armada 14 balloon was advanced without any resistance to the target lesion and inflated once when it ruptured at 12 atmospheres (atm).A second 3x200mm armada 14 balloon was advanced without resistance but when inflated once it ruptured at 14 atm.A non-abbott balloon (csi) was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 14 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15898596
Report Number2024168-2022-12103
Device Sequence Number5113802
Product Code DQY
UDI-Device Identifier08717648156113
UDI-Public08717648156113
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K102705
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model NumberA2030-200
Device Catalogue NumberA2030-200
Device Lot Number1080441
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/16/2022
Initial Report FDA Received Date12/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age80 YR
Patient SexMale
Patient Weight81 KG
-
-