| Model Number |
FR8A-TRL-A0, FR8A-TRL-B0 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Device Overstimulation of Tissue (1991); Perforation (2001)
|
| Date of Event |
11/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The clinical representative reported a trial procedure was aborted as the implanting clinician punctured into intrathecal space during first attempt with excess blood return.The clinician attempted to implant a level above.However, the patient felt overstimulated with testing and the clinician aborted the case.The patient was sent for mri immediately after the procedure was aborted.The patient is currently doing well and no further issues have been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The other adverse events issues questionnaire was reviewed for potential causes of the reported issue.Based on this review, the incident occurring before the implant of the device and implanting the device in an off-label location, have been ruled out as potential causes.However, the questionnaire shows medical intervention was required to prevent serious injury and tissue was punctured during the procedure.The stimulator is used to treat pain.The cause of the reported issue is incorrect surgical technique as the implanting clinician punctured into intrathecal space (user error - clinician).Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in other adverse events issues.Other adverse events issues rates remain acceptably low; thus, capa is not required.Other adverse events issues rates will continue to be tracked and trended.
|
| |
|