U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG., 6MM FEMORAL AIMER; ACCESSORIES, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG., 6MM FEMORAL AIMER; ACCESSORIES, ARTHROSCOPIC Back to Search Results
Model Number 234-020-046
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the device broke during the procedure.All pieces were retrieved.
 
Event or Problem Description
It was reported that the device broke during the procedure.All pieces were retrieved.
 
Additional Manufacturer Narrative
Alleged failure: when use of stryker femoral 6 mm aimer guide under knee arthroscopy operation, loose debri in the joint was seen coming from the guide - not from the k-wire.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root causes could be: 1) use of excessive force, 2) improper reprocessing cycles/agents used, 3) instrument not dried properly after reprocessing, 4) improper storage conditions.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG., 6MM FEMORAL AIMER
Common Device Name
ACCESSORIES, ARTHROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15898923
Report Number0002936485-2022-00719
Device Sequence Number14720502
Product Code NBH
UDI-Device Identifier07613154174836
UDI-Public07613154174836
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number234-020-046
Device Catalogue Number234-020-046
Device Lot Number767677
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/07/2022
Supplement Date Received by Manufacturer11/07/2022
Initial Report FDA Received Date12/01/2022
Supplement Report FDA Received Date01/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-