Model Number N/A
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event is recorded with zimmer biomet under (b)(4).Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that outside of surgery the cutter was used with a damaged comb.The cutter did not meet mesh test acceptance criteria due to an incomplete cut.There was no patient involvement.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further information is available at this time.
 
Additional Manufacturer Narrative
This event is recorded with zimmer biomet under: (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the cutter failed the mesh test with an incomplete cut.The cutter was returned as is.A definitive root cause cannot be determined.The event is confirmed.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Z.S.G.M. CUTTER 1:1 RATIO ( CAUTION: SHARP )
Common Device Name
EXPANDER, SURGICAL, SKIN GRAFT
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15900835
Report Number0001526350-2022-01161
Device Sequence Number20114432
Product Code FZW
UDI-Device Identifier00889024375437
UDI-Public(01)00889024375437(11)211221(10)652239219
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
PREAMENDMENT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00770301000
Device Lot Number65239219
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer03/13/2023
Initial Report FDA Received Date12/01/2022
Supplement Report FDA Received Date03/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/21/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-