Model Number CB030
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
Procedure performed: nephrectomy.Event description: [institution] report from the sales rep it was hard and difficult to open the scissors during the procedure.Completed the procedure by switching to another product.Initial investigation report the event unit was returned to us and visually inspected.There were no scratches or abnormalities on the blade (pic.1).The units will be returned to amr for further evaluation.Products available for return: 1 package, 1 scissors patient status: no patient injury.Intervention: completed the procedure by switching to another product.
 
Additional Manufacturer Narrative
The event unit is anticipated to return to applied medical.A follow up report will be provided following the completion of the investigation.
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Testing was performed on the event unit.However, the complainant¿s experience could not be replicated or confirmed as the device was able to open and close without difficulty.The event unit met current specifications and there were no visible non-conformances.Applied medical has reviewed the details surrounding the event and related products and is unable to determine the cause of the reported event or confirm that a product malfunction occurred.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
 
Event or Problem Description
Procedure performed: nephrectomy.Event description: [institution] report from the sales rep.It was hard and difficult to open the scissors during the procedure.Completed the procedure by switching to another product.Initial investigation report.The event unit was returned to us and visually inspected.There were no scratches or abnormalities on the blade (pic.1).The units will be returned to amr for further evaluation.Products available for return: 1 package, 1 scissors.Patient status: no patient injury.Intervention: completed the procedure by switching to another product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CB030, 5MMX35CM EPIX LAPA SCISSOR, 10/BX
Common Device Name
LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES)
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key15901621
Report Number2027111-2022-00840
Device Sequence Number5123040
Product Code HET
UDI-Device Identifier00607915110123
UDI-Public(01)00607915110123(17)241216(30)01(10)1438415
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062169
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/16/2024
Device Model NumberCB030
Device Catalogue Number101419701
Device Lot Number1438415
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2022
Supplement Date Received by Manufacturer11/21/2022
Initial Report FDA Received Date12/02/2022
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-