| Model Number |
B1050-100 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It is unknown if the device is returning for analysis.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that the 5.00x100mm armada 35 balloon dilatation catheter (bdc) was used in a procedure and a hub leak was noted; therefore the bdc was removed from the patient.There was no adverse patient effect and not clinically significant delay reported in the procedure.No additional information was provided.
|
| |
|
Event or Problem Description
|
|
Subsequent to the initial medwatch report, the following information was received: it was reported that the procedure was on (b)(6) 2022 and was to treat the superficial femoral artery.Another armada balloon was used to continue the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.Based on the information provided, a conclusive cause for the reported leak could not be determined.It may be possible that the inflation device was not properly connected to the sidearm or damage to the sidearm occurred; however, without having the device to examine, a conclusive cause could not be determined.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.B3, b6: date of event, date of tests.H3: device was not received.
|
| |
|